Why Do SCI and SCE Measurements Differ on the Same Sample?
Understanding SCI and SCE measurement differences
Why Do SCI and SCE Measurements Differ on the Same Sample?
It is normal for L*a*b* and ΔE results to differ when the same sample is measured in SCI and SCE modes. The main reason is that the two modes handle specular reflection from the sample surface differently. The difference is often more noticeable on glossy, shiny or textured samples.
Key takeaway: different SCI/SCE values do not mean one mode measured incorrectly. First confirm that both sides used the same mode and other measurement conditions.
SCI means Specular Component Included: specular reflection is included in the measurement. SCE means Specular Component Excluded: specular reflection is excluded. Because the reflected light included in each measurement differs, the resulting L*a*b* and ΔE may differ too. See our dedicated guide to what SCI/SCE mean and when to use each mode.
1. What does each mode measure differently?
SCI includes specular reflection and can reduce the influence of surface-gloss differences when comparing material color.
SCE excludes specular reflection, so surface shine, matte appearance and gloss differences are more likely to appear in the measurement result.
Neither mode measures the “right” versus “wrong” color. They handle surface reflection differently.
2. Why do glossy and matte samples often show larger differences?
When light hits a sample, some forms specular reflection related to its surface, while other light enters the material and is reflected more diffusely. If two samples have similar material color but one surface is shinier and the other more matte, SCI and SCE treat those reflected components differently. Their L*, a*, b* and ΔE can therefore differ.
There is no fixed difference size. More visible gloss may make SCI and SCE results diverge, but the actual difference also depends on material, color, surface structure, measurement aperture and instrument optics. A glossy sample does not always have a large difference.
3. How can L*a*b* change between SCI and SCE?
There is no single direction that applies to every sample. Since the modes handle specular reflection differently, L*, a* and b* may all be affected. L* lightness may change; a* may shift slightly along the red-green axis; b* may shift slightly along the yellow-blue axis; and ΔE may change as the L*a*b* values change. Compare SCI and SCE values and make sure both parties use the same mode.
Do not assume that SCE is always darker or always produces a larger ΔE. The result depends on the sample and comparison method.
4. Why can a supplier pass a sample while the customer rejects it?
If the same production sample yields different results, SCI/SCE is one of the first conditions to check. Also confirm that both sides use the same illuminant and observer; measurement aperture; master standard or a standard from the same source; sample location and orientation, especially for curved, textured or uneven samples; and ΔE formula.
5. Which samples are more likely to show SCI/SCE differences?
Glossy plastics: mold texture, gloss level and surface treatment can change readings.
Coatings and paint: gloss, coating smoothness and application condition directly affect surface reflection.
Printing and packaging: varnish, gloss or matte lamination, and paper type can affect appearance and measured material color.
Rubber and textured surfaces: roughness, embossing and surface structure change reflection.
6. Which mode should R&D, color matching and appearance inspection use?
Do not assume that R&D always needs SCI or quality control always needs SCE. First ask: are you comparing the material color itself, or the real appearance including surface gloss?
If you want to reduce the influence of surface gloss, consider SCI, for example when comparing formulations, raw materials or material color.
If you want the result to reflect surface appearance differences too, consider SCE, for example for finished-product appearance or gloss-versus-matte comparisons.
For more detail about the principles and use cases, see our dedicated SCI/SCE guide.
Frequently asked questions
Is it normal for SCI and SCE values to differ?
It can be normal. They handle specular reflection differently, so glossy samples may show different L*a*b* and ΔE.
Is SCE L* always lower than SCI L*?
Do not treat that as a rule. Results depend on sample color, gloss, surface structure and instrument optics.
Can ΔE values from SCI and SCE be compared directly?
It is not recommended to treat different modes as the same measurement condition. Choose one mode when establishing standards and acceptance limits.
What should customers and suppliers check first when values differ?
Confirm SCI/SCE, illuminant, observer, aperture, master standard, measurement location and color-difference formula.
Need an instrument with SCI/SCE capability?
Supported modes, apertures and color-data functions vary by model. If you are selecting equipment, see our spectrophotometer and instrument-selection guides. For small measurement areas, review models such as Aurora with different aperture configurations.
Customer and supplier SCI/SCE readings do not agree?
Share the measurement mode, instrument model, illuminant, aperture, master sample and sample condition so the measurement conditions can be reviewed.
Contact Fulcrum Technology | SCI/SCE differences | Spectrophotometers