LS127C Integrated UVC Irradiance Meter

The LS127C integrated UVC irradiance meter measures irradiance from UVC germicidal lamps. Its one-piece design includes a 1 m measuring hook for positioning the meter at the required location. Bluetooth supports remote viewing with a compatible app; actual range and compatibility depend on the environment, device and software. Confirm the source wavelength, measurement distance, illuminated area and range required by your procedure. Remote viewing does not replace UVC shielding, exposure controls or workplace safety procedures.

LS127C Integrated UVC Irradiance Meter

For field checks of UVC germicidal lamps, disinfection equipment and 253.7 nm mercury lamps

The LS127C integrated UVC irradiance meter measures irradiance and energy from UVC germicidal lamps, 253.7 nm / 254 nm mercury lamps, disinfection equipment, sterilization cabinets and related UV systems. Its one-piece design suits quick field measurements, spot checks and lamp intensity verification. If testing must follow a healthcare, disinfection-equipment or internal quality procedure, confirm the method against the applicable standard, measurement distance and site conditions.

01

Field measurements of UVC germicidal lamps

The LS127C is suited to irradiance checks of UVC germicidal lamps, disinfection equipment, sterilization cabinets and 253.7 nm mercury lamps. A measuring hook can help maintain the required measurement position and distance.

LS127C integrated UVC irradiance meter with measuring hook for germicidal lamp testing

A 1 m measuring hook is available for field checks of UVC germicidal lamps.

02

Measurement statistics at a glance

The meter displays real-time, maximum, minimum and average irradiance values, together with measurement duration. The energy display can also show accumulated energy for field inspection and record keeping.

LS127C UVC meter display showing real-time, maximum, minimum and average values
03

USB communication and computer software

A USB communication interface and dedicated computer software support measurement data review, organization and report preparation. Confirm the current software version and compatibility with Fulcrum before deployment.

LS127C UVC meter connected to computer software for measurement data management
04

Automatic shutdown and saved readings

The auto-off timer can be set from 1 to 255 minutes. Measurement data are saved when the meter switches off; after restart, the latest reading can be reviewed. Product information states storage for up to 9 records, suitable for short inspections and field logs.

LS127C UVC irradiance meter with automatic shutdown and storage for up to 9 records
05

Bluetooth data transfer for remote viewing

Bluetooth supports remote viewing of measurements with a compatible application. This can help when a UVC lamp is operating and it is not practical to view the instrument screen nearby. Actual range and compatibility depend on the environment, device and software. Continue to follow site safety procedures and confirm shielding and exposure controls.

Remote viewing of LS127C UVC irradiance readings over Bluetooth using a compatible application

Bluetooth viewing requires a compatible application; verify device and software requirements for your site.

Selection guidance

The LS127C is intended mainly for field checks of UVC germicidal lamps, 253.7 nm / 254 nm mercury lamps and disinfection equipment. For measurements inside equipment or deep enclosures, or when a separate probe is needed, consider the LS126C split-type UVC irradiance meter. For one instrument with interchangeable UVA, UVB and UVC probes, consider the LS125 UV irradiance meter. Select based on source wavelength, measurement distance, illuminated area, measurement range and site safety procedure.

Contact Fulcrum about selection

Reference specifications from the available product information. Confirm the current configuration and acceptance criteria with Fulcrum before purchase or use.

ParameterSpecification
Spectral response230–280 nm; peak wavelength λp = 254 nm
Irradiance range0–20,000 μW/cm²
Irradiance resolution0.1 μW/cm²
Energy range0–9,999,999 μJ/cm²
Relative indication error (referenced to NIM standard)H < 50 μW/cm²: ±4 μW/cm²
H ≥ 50 μW/cm²: ±8%H
Long-wavelength response error<60%
Cosine response error10%
Linearity error±1%
Short-term stability±1% after 30 minutes powered on
Fatigue characteristicAttenuation <2%
Zero error±0.1% of full scale
Response time<0.5 s
Recording duration0–99,999 s
Selectable unitsμW/cm² (default), mW/cm², W/m²
Bluetooth rangeAbout 50 m in an open area
Instrument dimensions188.5 × 75.2 × 30.3 mm
WeightAbout 232 g including batteries
Display240 × 160 dot-matrix LCD
Power2 AA alkaline batteries
Operating environment0–40°C; humidity <85% RH

Specifications may change. Contact Fulcrum for the latest information.


LS127C Operation Video

UVC Irradiance Measurement Demonstration

Watch the video for an introduction to operation and on-site measurement with the LS127C integrated UVC UV irradiance meter. The LS127C is suitable for on-site inspection of UVC germicidal lamps, 253.7/254 nm mercury lamps, disinfection equipment and sterilization cabinets. Confirm the source wavelength, measurement distance, illuminated area, measurement range and site safety procedures before testing.

top