Large 20 mm Measurement Aperture
Covers a larger sampling area and suits textured, granular or unevenly colored samples, helping reduce variation from single-point sampling.
The NR20XE is a 45°/0° upright colorimeter with a Φ20 mm aperture for comparing samples with texture, uneven color distribution or a larger sampling area. Use it for incoming inspection, in-process checks, routine quality control and pre-shipment color review. Its upright design may be evaluated for damp surfaces or selected liquid-related applications. Confirm sample material, container, surface condition, moisture, positioning and quality-control workflow before use.
Designed for large-area sampling, fast color comparison, on-site quality checks and computer-based color data management.
Covers a larger sampling area and suits textured, granular or unevenly colored samples, helping reduce variation from single-point sampling.
Depending on sample conditions, evaluate damp-surface or liquid-related measurement. Confirm the container, liquid-surface stability and measurement setup first.
Helps align the instrument with the measurement point for fixed-location checks, repeat measurements and quick on-site decisions.
The 3nh optical-path and dynamic-integration design can be considered when assessing stability and repeatability; actual results depend on the sample and method.
Useful to evaluate directional, polarized or surface-reflective materials and help limit the effect of measurement angle.
Approximates common visual observation and supports surface color comparison with CIE Lab and ΔE evaluation.
Supports mobile inspection, field measurement and production-line sampling while reducing disposable battery replacement.
Connect to a computer for data review, quality records and color-difference analysis. Functions depend on software version and workflow.
Compare color across samples with patterns, granules, fibers or uneven color distribution using a large aperture.
Evaluate damp surfaces or liquid-related measurement by sample conditions; confirm how the sample will be held and measured.
Use for incoming inspection, in-process sampling, batch comparison and pre-shipment color checks.
Use CQCS3 to review data and maintain quality records as part of a color-management workflow.
The NR20XE is intended for 20 mm large-aperture color comparison, textured or damp surfaces and routine quality control. For glossy, transparent, translucent, curved or highly uneven samples, first confirm aperture, sampling position, sample fixture, averaging method and internal acceptance criteria. If full spectral data, multiple illuminants, SCI/SCE switching or supply-chain color traceability is required, consider a handheld spectrophotometer.
Use these specifications as a reference when evaluating, purchasing and integrating the NR20XE high-precision colorimeter into routine quality control. It uses 45/0 measurement geometry and a large Φ20 mm aperture for larger samples, textured or damp surfaces, and color comparison. Confirm the sample material, surface condition, measurement area, color uniformity and quality-control workflow before implementation.
| Model | NR20XE colorimeter |
|---|---|
| Measurement geometry | 45/0 (45° illumination, 0° viewing); complies with CIE No. 15 and GB/T 3978 |
| Detector | Silicon photodiode array |
| Spectral method | Not applicable |
| Integrating sphere size | Φ40 mm |
| Illumination source | Full-spectrum LED |
| Wavelength range | Not applicable |
| Illuminant | D65 |
| Wavelength interval | Not applicable |
| Reflectance measurement range | Not applicable |
| Measurement aperture | Φ20 mm |
| Measurement mode | SCI |
| Color spaces | CIE Lab, XYZ, LCh |
| Color difference formula | ΔE*ab |
| Other colorimetric indices | Not applicable |
| Observer angle | 10° |
| Displayed parameters | Sample color values, color difference, pass/fail result and color tendency |
| Measurement time | Approximately 1.5 seconds |
| Repeatability | ΔE*ab within 0.08 (after warm-up and calibration; average of 30 white-tile measurements at 5-second intervals) |
| Inter-instrument agreement | ΔE*ab within 0.4 (average measurement of 12 BCRA Series II color tiles) |
| Measurement method | Single measurement |
| Positioning | Alignment positioning |
| Dimensions | 205 × 67 × 80 mm |
| Weight | Approximately 400 g |
| Battery | Rechargeable lithium battery |
| Illumination lamp life | Approximately 5 years or more than 1.6 million measurements |
| Display | 2.8-inch TFT full-color display |
| Connection | USB |
| Data storage | 100 reference samples; 20,000 test samples |
| Software support | CQCS3 color quality management software |
| Languages | Traditional Chinese, Simplified Chinese and English |
| Operating conditions | 0–40°C; 0–85% relative humidity, non-condensing |
| Storage temperature | −20–50°C, non-condensing |
| Standard accessories | Power adapter, manual, wrist strap, white tile, USB cable, downloadable color quality management software and Φ20 platform measurement aperture |
| Optional accessories | Mini printer and powder test box |
| Notes | Specifications may change. Confirm the latest information with Fulcrum Technology or the manufacturer. |
Selection note: The NR20XE is suited to large-aperture 20 mm color comparison, textured or damp surfaces and routine quality control. For glossy, transparent, translucent, curved or highly uneven samples, first confirm the aperture, sampling position, fixture, averaging method and internal acceptance criteria. For full spectral data, multiple illuminants, SCI/SCE switching or supply-chain color traceability, consider a spectrophotometer.