LS126C UVC Split-Type UV Irradiance Meter

The LS126C split-type UVC UV irradiance meter measures irradiance and energy from UVC germicidal lamps, 253.7 nm mercury lamps, disinfection equipment and sterilization chambers. Place the remote probe at the measurement point while keeping the meter outside the equipment, making it suitable for enclosures, equipment interiors and confined spaces. Confirm the wavelength, working distance, illuminated area, range and site safety or quality-control procedure when selecting a configuration.

LS126C UVC Split-Type UV Irradiance Meter

For UVC germicidal lamps, disinfection equipment and 253.7 nm mercury lamps

The LS126C split-type UVC UV irradiance meter measures irradiance and energy from UVC germicidal lamps, 253.7 nm mercury lamps, disinfection equipment, sterilization chambers and related UV systems. Position the separate probe at the test point while keeping the meter outside the equipment. This design is useful for measurements inside equipment, enclosures and confined spaces where the display is difficult to view directly. Confirm the source wavelength, measurement distance, illuminated area, range and quality-control procedure for each application.

01

UVC germicidal lamp measurement

Measure irradiance and accumulated energy from 253.7 nm UVC germicidal lamps, disinfection equipment and related applications in healthcare, pharmaceutical production and environmental monitoring. Confirm the source wavelength, measurement distance and equipment layout before testing.

LS126C split-type UVC irradiance meter with test hook for germicidal lamp measurements

A test hook can be used for UVC germicidal lamp measurements when suitable for the site setup.

02

Measurement statistics

The display shows real-time power, measurement statistics, elapsed time and test date and time to help operators document UVC source conditions and results.

LS126C UVC irradiance meter displaying real-time power, measurement time and statistics
03

USB connection and PC software

Connect the meter to a computer through USB and use the PC software to view and organize measurement data and prepare reports. Confirm data-format and reporting requirements with your quality-control workflow.

LS126C UVC irradiance meter using USB and PC software for data management
04

Automatic shutoff and data storage

Set automatic shutoff from 1 to 10 minutes. Measurement data is retained when the meter turns off, and the latest result can be viewed after restart. The meter stores up to 25 sets of readings for inspections and short-term testing.

LS126C UVC irradiance meter supporting automatic shutoff and storage of up to 25 measurement sets
05

Bluetooth data transfer for remote viewing

Bluetooth data transfer can support remote viewing through a compatible WeChat mini program. This can help operators monitor readings without standing beside an operating UVC lamp. Actual availability depends on the device, application and site environment. Remote viewing does not replace shielding, exposure controls or workplace safety procedures.

LS126C UVC irradiance meter viewed remotely through Bluetooth and a compatible WeChat mini program

Remote data viewing requires a compatible WeChat mini program and device. Confirm availability for your setup.

Choosing a meter

The LS126C is intended for UVC germicidal lamps and 253.7 nm mercury lamps. For UVA LED curing sources or UVA measurements in confined spaces, see the LS126A UVA irradiance meter. For one meter with interchangeable UVA, UVB and UVC probes, see the LS125 UV irradiance meter with interchangeable probes. Confirm wavelength, distance, illuminated area, range and quality-control requirements before selecting a configuration.

Contact Fulcrum about configuration

The following specifications are for reference. Confirm the measurement wavelength, distance, illuminated area, range and applicable quality requirements for your test setup.

ItemSpecification
Spectral response230–280 nm; peak wavelength λp = 254 nm
Irradiance range0–20,000 μW/cm²
Irradiance resolution0.1 μW/cm²
Energy range0–9,999,999 μJ/cm²
Relative indication errorCompared with NIM: H < 50 μW/cm², ±4 μW/cm²; H ≥ 50 μW/cm², ±8%H
Long-wavelength response error<60%
Cosine response error10%
Linearity±1%
Short-term stability±1% after 30 minutes
Fatigue error<2%
Zero drift±0.1% FS
Sampling rate6 readings/s
Probe apertureΦ10 mm
Power-data intervalSelectable: 1, 10 or 60 seconds
Logging duration10, 100 or 600 minutes
Bluetooth rangeApproximately 50 m in an unobstructed area
Probe dimensions39 × 15 mm
Probe cable1 m
Meter dimensions148 × 76 × 26 mm
WeightApproximately 293 g, including batteries
Display240 × 160 LCD
Power supply4 × AAA alkaline batteries
Operating environment0–40°C; <85% RH

Specifications may be updated by the manufacturer. Please confirm the latest configuration with Fulcrum before purchase or testing.


LS126C Operation Video

UVC Split-Type Irradiance Meter Measurement Demonstration

This video provides an overview of LS126C operation and on-site measurement. The separate probe can be positioned at the measurement point while the meter remains outside the enclosure. LS126C is suitable for UVC germicidal lamps, 253.7 nm mercury lamps, disinfection equipment and sterilization cabinets. For actual testing, confirm the source wavelength, measurement distance, illuminated area, measurement range and site safety procedures.