LS129 UV Irradiance Meter Module
LS129 UV Irradiance Meter Module
UV Monitoring for Equipment Integration and Online Measurement
The LS129 is a digital UV probe module designed to integrate UV intensity and energy monitoring into equipment, production lines and automation systems. It uses RS485 communication with the MODBUS protocol. The detector and controller are separate for flexible installation and can connect to a PC, PLC or HMI. Typical applications include UV curing equipment, UV LED sources, UVC germicidal equipment and automated process monitoring. Confirm the spectrum, irradiance range, mounting space, communication needs and control system before selection.
LS129 Communication Parameters
MODBUS address 1 is the default and can be changed with the commissioning tool or protocol settings. Adjustable address range: 1–247.
Default baud rate: 9600 bps. Selectable rates: 4800, 9600, 19200 or 38400 bps.
Read measurement data and configure the address and baud rate for PLC, PC or HMI integration.
Probe Wiring Reference
Wire colors may vary by production batch. Verify the supplied documentation or confirm with Fulcrum before wiring; do not rely on color alone.
| Wire color | Function |
|---|---|
| Black (with heat-shrink sleeve) | Shield |
| Red | Power positive (DC 5–24 V) |
| Black | Power negative |
| Green / blue | RS485− |
| Yellow / white | RS485+ |
Optional Commissioning Tools
The following tools are optional. Select them according to the integration method, control system and installation conditions, and confirm the quoted configuration.
The LS129 is a UV irradiance meter module for equipment integration and online monitoring, not a general-purpose handheld inspection meter. For spot checks or field inspections, consider the LS125 multi-probe UV irradiance meter. For PLC, HMI or host-computer integration, assess protocol, power, mounting space and source spectrum. Contact Fulcrum for selection and system-integration guidance.
The LS129 supports RS485 and MODBUS and can be tested with commissioning tools or PC software. Contact Fulcrum to confirm the current software, protocol and wiring documentation version.
View LS129 software and support resourcesLS129 Model Comparison
Choose the detector version according to the UV source spectrum and required measurement range. Specifications below are reference values; confirm the ordered version and current configuration before integration.
| Parameter | LS129-UVALED | LS129-UVA | LS129-UVC |
|---|---|---|---|
| Spectral response | 340–420 nm; calibrated at 395 nm | 315–400 nm; peak wavelength 365 nm | 230–280 nm; peak wavelength 254 nm |
| Typical application | UV LED curing and 365–405 nm LED source monitoring | UVA mercury lamps, exposure, aging and general UVA monitoring | Germicidal lamps, 254 nm mercury lamps and disinfection equipment |
| Irradiance range | 0–20,000 mW/cm² | 0–2,000 mW/cm² | 0–200,000 μW/cm² |
| Energy range | 0–4×10⁹ mJ/cm² | 0–4×10⁹ mJ/cm² | 0–4×10⁹ μJ/cm² |
| Resolution | 1 mW/cm² / 0.1 mW/cm² | 0.1 mW/cm² | 0.1 μW/cm² |
| Accuracy | H < 50: ±5 mW/cm²; H ≥ 50: ±10%H | H < 5: ±0.5 mW/cm²; H ≥ 5: ±10%H | H < 50: ±5 μW/cm²; H ≥ 50: ±10%H |
| Sampling rate | 2048/s | 2048/s | 6/s |
| Data refresh | 200 ms | 200 ms | 200 ms |
| Aperture | Φ10 mm | Φ10 mm | Φ10 mm |
| Communication cable | 1 m; rated to 80°C | 1 m; rated to 80°C | 1 m; rated to 80°C |
| Detector cable | 0.5 m; rated to 200°C | 0.5 m; rated to 200°C | Not applicable |
| Interface / protocol | RS485 / MODBUS | RS485 / MODBUS | RS485 / MODBUS |
| Power supply | DC 5–24 V, 20 mA | DC 5–24 V, 20 mA | DC 5–24 V, 20 mA |
| Operating environment | −25 to 85°C; <85% RH, non-condensing | −25 to 85°C; <85% RH, non-condensing | −25 to 85°C; <85% RH, non-condensing |
| Controller dimensions | 72.5×32×15.1 mm | 72.5×32×15.1 mm | 72.5×32×15.1 mm |
| Detector dimensions | 35×15×14.5 mm | 35×15×14.5 mm | 35×15×14.5 mm |
H denotes the measured irradiance value. Confirm the source spectrum, range, calibration requirement, environmental conditions and product revision with Fulcrum before final selection.
Installation and Integration Overview
The LS129 is designed to be incorporated into UV equipment and production systems. Before installation, confirm the detector version matches the source spectrum, provide a stable DC 5–24 V supply, and plan the detector position, cable route and controller mounting location.
Commissioning Checklist
- Verify the ordered model (UVALED, UVA or UVC) and the applicable irradiance range.
- Check the wiring against the documentation supplied with the current production batch; wire colors can vary.
- Set a unique MODBUS address and a supported baud rate, then confirm communication with the PLC, PC or HMI.
- Compare readings with the intended operating conditions and establish the application's own alarm or process limits.
For spot measurements and field inspections, see the LS125 multi-probe UV irradiance meter. For system selection and integration questions, contact Fulcrum.
Software and Technical Support
Use the LS129 software and support resources for available setup materials. Confirm the current software and document revision with Fulcrum before commissioning. Final system design, alarm logic and connection to a database or host platform may require application engineering.