CHNSpec SPF-600 / SPF-620 / SPF-660 Sunscreen Protection Factor Analyzer

CHNSpec SPF-600 / SPF-620 / SPF-660 sunscreen protection factor analyzers are designed for in-vitro sunscreen performance analysis. They measure SPF, UVAPF, UVA/UVB (UVR), critical wavelength, spectral transmittance, and absorbance. The three models differ mainly in measurement wavelength range and analytical capability: SPF-600 covers 280–400 nm, SPF-620 extends to 450 nm, and SPF-660 covers 280–500 nm with SPF 1–2000 measurement and HEVIS high-energy visible-light analysis.

SPF-600 / SPF-620 / SPF-660 Sunscreen Protection Factor Analyzers

The CHNSpec SPF-600, SPF-620, and SPF-660 are in-vitro instruments for sunscreen performance and spectral analysis. They measure SPF, UVAPF, UVA/UVB (UVR), critical wavelength, spectral transmittance, and absorbance. Choose a model by wavelength range, SPF range, supported methods, and the need for HEVIS analysis.

ModelWavelength rangeSPF rangeKey differences
SPF-600280–400 nm1–1000For general SPF, UVAPF, and UVA/UVB sunscreen performance analysis.
SPF-620280–450 nm1–1000Extends to 450 nm and adds methods including ISO 23675 and FDA 2012 & 2019.
SPF-660280–500 nm1–2000Extends to 500 nm and adds HEVIS high-energy visible-light analysis.

Model selection: Consider SPF-600 for general SPF/UVAPF analysis; SPF-620 when the 450 nm range and additional methods are required; or SPF-660 for 500 nm coverage, SPF up to 2000, or HEVIS analysis.

Full-spectrum scan in approximately one second

High-speed spectral acquisition and a detector array capture the full wavelength range in about one second per scan. This can reduce the time required for point-by-point scanning and help improve sunscreen formulation testing and comparison efficiency.

Integrated operation and real-time analysis

The system integrates a processor, embedded operating system, and touchscreen interface. It supports spectral acquisition, dark-current correction, and calculation of SPF and UVA/UVB results, allowing the basic test workflow to run without an external computer.

High dynamic range; SPF-660 measures up to SPF 2000

The manufacturer's high-dynamic-range photoelectric detection system is designed to handle signal changes from high- to low-transmittance samples. SPF-600 and SPF-620 cover SPF 1–1000; SPF-660 covers SPF 1–2000.

Broad wavelength coverage up to 500 nm

Available ranges depend on the model: 280–400 nm, 280–450 nm, or 280–500 nm. SPF-660 extends beyond the UVA and UVB bands to 500 nm for HEVIS high-energy visible-light analysis.

Multiple sunscreen test methods

SPF-600 supports the Cosmetics Safety Technical Specification, ISO 24443, COLIPA (2011), and Boots Star Rating (2011). SPF-620 and SPF-660 additionally support ISO 23675 and FDA 2012 & 2019 sunscreen labeling and efficacy methods.

Select test methods according to product-development objectives, target markets, and applicable requirements. In-vitro analysis should not be interpreted as replacing every human test or regulatory procedure.

Need help choosing the right analyzer?

Share your sunscreen product type, measurement objectives, required test methods, and wavelength range. Fulcrum Technology can help assess whether the SPF-600, SPF-620, or SPF-660 fits your application.

Discuss sample suitability and request a quotation

SPF-600 / SPF-620 / SPF-660 Series — Specifications

SpecificationSPF-600SPF-620SPF-660
Wavelength range280–400 nm280–450 nm280–500 nm
Wavelength accuracy±0.1 nm
Spectral resolution (FWHM)< 2 nm
Wavelength interval / data spacing1 nm
Optical geometryD/0 (integrating-sphere illumination, 0° field of view)
Integrating sphere materialFPET
Integrating sphere port diameter1.27 cm
Sample illumination area1.26 cm²
Light sourcePulsed xenon lamp
UV irradiation dose per test< 0.2 J/cm²
Sample stageManual three-dimensional positioning stage
Transmittance range0–100%
Absorbance range0–3.3 A
SPF measurement range1–10001–10001–2000
Single-scan timeApproximately 1 s
Measured parametersSPF, UVAPF, UVA/UVB (UVR), critical wavelength, spectral transmittance and absorbanceSame as SPF-600Same as SPF-600, plus HEVIS
Supported methodsCosmetics Safety Technical Specification; ISO 24443; COLIPA (2011); Boots Star Rating (2011)Cosmetics Safety Technical Specification; ISO 23675; ISO 24443; FDA 2012 & 2019; COLIPA (2011); Boots Star Rating (2011)Same as SPF-620
Computer interfacesUSB 2.0, RS-232
Power12 V / 5 A
Operating conditions0–50°C; 0–70% RH, non-condensing
Dimensions with operating stand523 (D) × 332 (W) × 290 (H) mm
Dimensions without operating stand325 (D) × 332 (W) × 290 (H) mm
Optional accessoriesAS01 PMMA reference plate setAS01 PMMA reference plate set; AS02 PMMA test plate setAS01 PMMA reference plate set; AS02 PMMA test plate set

Specifications are compiled from CHNSpec product information for the SPF-660 series. Confirm the exact model, supplied configuration and accessories when ordering.

top