MS3008 Eight-Angle Multi-Angle Spectrophotometer
MS3008 Automotive Multi-Angle Spectrophotometer | Eight-Angle Measurement for Metallic, Pearlescent, and Special-Effect Finishes
MS3008 Automotive Multi-Angle Spectrophotometer is designed for applications that require multi-angle evaluation of metallic, pearlescent, and other special-effect colors, including automotive topcoats, plastic exterior parts, and industrial coatings. Its six light sources and two receivers support measurements at eight angles, helping document color differences under different viewing geometries.
For effect coatings, metallic or pearlescent pigments, and specialty finishes, appearance can vary with viewing angle, illumination, surface curvature, and sample condition. The MS3008 supports evaluation of multi-angle color difference, sparkle, color-flop, and graininess. Confirm suitability against the sample, required angles and effect parameters, and your quality-control workflow.
MS3008 Multi-Angle Spectrophotometer Features
Designed for multi-angle color and effect-parameter evaluation of automotive coatings, special-effect colors, metallic and pearlescent finishes, and curved exterior parts.
Eight-angle measurement
Six light sources and two receivers enable measurements at eight angles, helping build color data for metallic, pearlescent, and special-effect finishes across viewing geometries.
Multi-angle result display
The touchscreen displays results from different angles for convenient review of multi-angle color differences, sample color values, and pass/fail decisions.
Effect-parameter evaluation
Evaluate effect parameters such as sparkle, color-flop, and graininess for quality-control workflows that need to distinguish special-effect surface appearance.
256-pixel dual-array CMOS sensor
A 256-pixel dual-array CMOS image sensor provides the sensing basis for multi-angle spectral data collection and color analysis.
Blue-enhanced full-spectrum LED source
The blue-enhanced full-spectrum LED source supports evaluation of spectral distribution and color measurements across the visible range.
Concave-grating spectroscopic technology
Concave-grating spectroscopy supports spectral data analysis and multi-angle color-data generation for applications that track color variation between batches.
Reference tile and calibration
Use the reference white tile for routine calibration and to maintain measurement conditions. Confirm calibration frequency and procedure against the manufacturer instructions and your quality-control workflow.
MCU and wireless connectivity
An industrial real-time MCU supports Wi-Fi and Bluetooth 5.0. Choose a deployment approach based on on-site measurement and data-management needs.
Handheld operation
The body shape and measurement-button placement are designed for handheld operation in laboratories, on-site inspections, and exterior-part measurements.
Color-camera preview and positioning
The built-in color camera helps operators confirm the target area, supporting repeatable positioning on fixed measurement points, curved parts, and other samples.
Multiple color spaces and illuminants
Supports color spaces including CIE LAB, XYZ, Yxy, LCh, βxy, and DIN Lab99, as well as illuminants such as D65, A, C, D50, D55, D75, and the F series.
Measurement data display and analysis
The screen displays spectra and data, sample color values, color differences, pass/fail results, color simulation, and effect values for convenient on-site review and quality-control records.
Pre-purchase considerations
The MS3008 is intended for eight-angle color measurement and special-effect surface evaluation. For samples with metallic or pearlescent pigments, curved parts, or special-effect coatings, first confirm the required angles and effect parameters, reference-color method, sample condition, and internal acceptance criteria against your quality-control process.
Contact Fulcrum Technology for MS3008 selection adviceMS3008 Automotive Multi-Angle Spectrophotometer Specifications
The specifications below support product selection, purchasing, and quality-control evaluation. The MS3008 provides eight measurement angles for metallic, pearlescent, and special-effect colors, automotive topcoats, plastic exterior parts, and industrial coatings. Confirm the final configuration based on the sample, pigments, surface gloss, required angles and effect parameters, and your quality-control workflow.
| Model | MS3008 series |
|---|---|
| Measurement geometry | Eight measurement angles (six light sources and two receivers) |
| Color measurement angles | 45° receiver: 45as-15°, 45as15°, 45as25°, 45as45°, 45as75°, 45as110°; 15° receiver: 15as-45°, 15as-15° |
| Standards | ASTM D2244, ASTM E308, ASTM E1164, ASTM E2194, ASTM E2539, DIN 5033, DIN 5036, DIN 6174, DIN 6175-1, DIN 6175-2, ISO 7724, ISO 11664-4, SAE J1545 |
| Applications and capabilities | For multi-angle color measurement and quality-control comparison of metallic, pearlescent, and other complex special-effect finishes. Confirm evaluation against the sample, reference-color setup, and internal acceptance criteria. |
| Light source | White full-spectrum LED with blue enhancement |
| Light-source life | 5 years or 3 million measurements |
| Spectroscopic method | Concave grating |
| Detector | 256-pixel dual-array CMOS image sensor |
| Wavelength range | 400–700 nm |
| Wavelength interval | 10 nm |
| Reflectance range | 0–600% |
| Spectral bandwidth | 10 nm |
| Measurement aperture | Illumination Φ23 mm / sampling 9 × 12 mm; custom option: illumination Φ10 mm / sampling 6 × 8 mm |
| Color spaces | CIE LAB, XYZ, Yxy, LCh, βxy, DIN Lab99 |
| Color-difference formulas | ΔE*ab, ΔE*94, ΔE*cmc (2:1), ΔE*cmc (1:1), ΔE*00, DIN ΔE99, DIN 6175, and automotive-brand color-difference formulas |
| Additional colorimetric indices | Flop Index; radiance value |
| Observer angle | 2° / 10° |
| Illuminants | D65, A, C, D50, D55, D75, F1, F2 (CWF), F3, F4, F5, F6, F7 (DLF), F8, F9, F10 (TPL5), F11 (TL84), F12 (TL83 / U30) |
| Display | Spectral graph/data, sample color values, color difference/graph, pass/fail results, color simulation, sample effect values, and effect differences |
| Measurement time | About 1 second per angle and about 6 seconds for all angles (effect-color testing excluded) |
| Color repeatability | Spectral reflectance: standard deviation within 0.08%; colorimetric value: 0.03 ΔE*ab (average of 30 white-tile readings at 5-second intervals after warm-up and calibration) |
| Color reproducibility | ΔE* < 0.10 (gray BCRA tile average); ΔE* < 0.25 (color BCRA tile average) |
| Instrument-to-instrument difference | 0.18 ΔE*00 (average across 12 BCRA Series II color tiles) |
| Effect parameters | Sparkle, color flop, and graininess |
| Effect measurement | Six-angle sparkle and color-flop data: 15as-45°, 15as-30°, 15as-15°, 15as15°, 15as45°, 15as80°; 15d diffuse graininess |
| Effect repeatability | Short-term sparkle repeatability: 0.12% (standard deviation across 10 measurements; average of 10 color-tile readings at 10-second intervals after warm-up and calibration); short-term graininess repeatability: 0.09% (same method) |
| Effect reproducibility | Sparkle reproducibility: 1.9% (standard deviation across 10 measurements; average across 12 BCRA Series II tiles); graininess reproducibility: 1.4% (same method) |
| Trigger modes | Pressure-sensor, button, and software trigger |
| Measurement modes | Single, average (1–99 readings), and continuous (1–99 readings) measurement |
| Positioning | CCD color preview |
| Dimensions | 195 × 83 × 128 mm (L × W × H) |
| Weight | Approx. 1 kg |
| Battery | Lithium battery, 3.7 V, 3200 mAh; up to 6,000 continuous tests within 8 hours on a full charge |
| Display | 3.5-inch full-color TFT capacitive touchscreen |
| Interface | USB, Bluetooth 5.0 |
| Data storage | 1,000 reference samples and 4,000 test samples |
| Languages | Chinese (Simplified and Traditional), English |
| Operating temperature | 10–50°C; humidity up to 85%, non-condensing |
| Storage temperature | -20–50°C; humidity up to 85%, non-condensing |
| Calibration | Built-in white-tile parameters; external white tile, black trap, and color tiles |
| Calibration interval | 4, 8, or 24 hours; or calibration at startup |
| Standard accessories | Power adapter, USB cable, manual, quality-management software (download from official website), calibration box, black trap, protective cover, and wrist strap |
| Optional accessory | Mini printer |
| Note | Specifications are for reference only; the actual product supplied governs. |
Selection note: The MS3008 supports eight-angle color measurement and evaluation of sparkle, color flop, graininess, and other special-effect surfaces. For metallic or pearlescent pigments, curved parts, or special-effect coatings, confirm the measurement angles, required effect parameters, reference-color setup, sample condition, and internal acceptance criteria. Verify supplied items, software functions, accessories, and measurement conditions with Fulcrum Technology.